Nitrogen Detector for GC 

7090

 

Using the accuracy and precision of the PAC-Antek® Pyro chemiluminescence technology, the GC detector 7090NS nitrogen / sulfur provides an equimolar response for all nitrogen and sulfur compounds. Chemiluminescence is the clean, rapid, without interference method to determine all of the nitrogen and sulfur compounds in a variety of matrices.

Advantages:

    • Simultaneous detection of sulfur and nitrogen
    • Parallel use of a FID detector or other, for 3 results by injection
    • Equimolar response for any type of nitrogen and sulfur product
    • pico-gram sensitivity

 

Test methods : ASTM D5504, ASTM D5623, GPA 2199-99

 

Applications : GC detector

 

 

Technical specificationsOptionsDownload the brochure
Detection Method: Pyro-chemiluminescence
Sensitivity:  <3 pgN/sec or <1 pgS/sec (individually optimized) ;

<30 pgN/sec et <10 pgS/sec (simultaneous detection)

Selectivity >107  to hydrocarbon and nitrogen/sulfur by weight
Linearity: >104
Equimolarity: <±10%
Gas : 'GC Grade' or 'Zero Grade', <5ppm H2O,

Dry oxygene: 2–20 cc/min 99.75%, 3.0 bar (40 psig),

Dry hydrogene: 50–400 cc/min 99.995%, 3.0 bar (40 psig)

Outputs: 1 mV, 1 V, 10 V analog
Combustion Zone Temperature: 950ºC–1100ºC by 1ºC increments
Power: 115VAC 50/60 Hz 12A or 230 VAC 50/60Hz 6A
Dimensions: 36 cm width (14') x 59 cm length (23') x  61 cm height (24')
Weight: 28 kg (62 lb)

 

  • Nitrogen detection kit: in addition to the Model 7090S sulfur detector.

 

  • Sulfur detection kit: in addition to the Model 7090N nitrogen detector.

 

  • Vacuum pump: low maintenance, high capacity. Without oil for more security for applications with oxygen.

 

  • Mounting Kit: includes all the specific parts for mounting on a GC.

 

 